When used with a current clamp, this standalone data logger measures and stores up to 127,232 a.c. and d.c. current readings over a 0 to 1000 amp d.c. measurement range (0 to 723 amps for a.c.). In ‘energy monitoring’ mode, this data is converted into power, energy (using a user defined voltage value) and cost (using a user supplied energy unit cost). The millivolt measurement mode allows for direct measurement of voltage, up to 1V d.c. (700mV a.c.).
The user can easily set up the logger and view downloaded data by plugging the data logger into a PC’s USB port and using the supplied software. The software has provision for a clamp scaling factor (the clamp input-output ratio of amps to millivolts). Stored data can then be graphed, printed and exported to other applications. The high contrast LCD can show a variety of current, power, energy and cost information. At the touch of a button, the user can cycle between the most recent, maximum and minimum measurement values. The data logger is fitted with 2 replaceable AA batteries, which can allow logging for up to 6 months.
Compatible with a.c. and d.c. current clamps
Energy monitoring mode (to calculate power, cumulative energy and cumulative cost)
AC and DC millivolt measurement mode
Logging rates between 1s and 12hr
4mm banana plug inputs
|
Specification
|
Min.
|
Typ.
|
Max.
|
Unit
|
|
Measurement range (a.c.)
|
0
|
|
723
|
mV a.c. RMS
|
|
Equivalent clamp measurement range (a.c.)*
|
0
|
|
723
|
A a.c.
|
|
Accuracy (a.c.)
|
|
5
|
|
%
|
|
Frequency
|
|
|
500
|
Hz
|
|
Measurement range (d.c.)
|
0
|
|
1000
|
mV d.c.
|
|
Equivalent clamp measurement range (d.c.)*
|
0
|
|
1000
|
A d.c.
|
|
Accuracy (d.c.)
|
|
2
|
|
%
|
|
Measurement resolution
|
|
250
|
|
μV
|
|
Equivalent resolution*
|
|
250
|
|
mA
|
|
Display resolution**
|
1
|
|
|
mA/mV
|
|
Logging rate (Current measurement)
|
Every 1s
|
|
Every 12hr
|
-
|
|
Logging rate (Energy measurement)
|
Every 1s
|
|
Every 10s
|
-
|
|
Operating temperature range
|
-10 (-31)
|
|
+50 (176)
|
ºC (ºF)
|
|
Battery life***
|
|
|
6
|
Months
|
* The standard clamp output ratio is 1mV per amp
** Display resolution will change, depending on the number of digits in use. The display will auto range to make the best use of available space